The differential amplifier and polarity
Clinical neurophysiology does not measure the voltage of a single point. The amplifier magnifies the difference between two inputs: Input 1 − Input 2. Anything identical at both inputs, such as mains interference on the whole body, ideally disappears.
How well the amplifier suppresses common signals is expressed as the common-mode rejection ratio (CMRR), in decibels. The higher the value, the better the rejection; but this advantage only reaches real life when the impedances of the two inputs are balanced.
Deflection direction is a convention
The neurophysiological convention is that when Input 1 is more negative than Input 2 the trace deflects upward. Most EEG and many evoked potential recordings follow this rule, but some devices and publications display it the other way, so check the polarity marker on the page.
A wave's name describes its true polarity, not its direction on the screen. For example, VEP traces are drawn positive-up in most laboratories, while SEP traces are often drawn negative-up. If the cables are swapped the same wave inverts, but its physiology does not change.