The electrode: from ionic current in tissue to electrons in the wire
In the body current is carried by ions; in the wire, by electrons. The electrode is the boundary between these two worlds, and that boundary is not perfect: a voltage arises spontaneously between metal and electrolyte. This is called the half-cell potential, and it can be far larger than the EEG signal we want to measure.
If two electrodes are made of the same material and are in the same condition, these voltages largely cancel. Problems begin when the electrodes differ or the contact moves: the voltage fluctuates and the recording shows slow drift or sudden jumps.
Why does impedance matter?
Impedance is the total opposition to alternating current. A high value alone is less harmful than inequality between electrodes, because mains interference then divides unequally between the two inputs and the amplifier magnifies that difference as if it were a real signal.
- Cleaning oil and dead skin lowers contact impedance.
- Silver/silver chloride (Ag/AgCl) electrodes are more stable for slow signals and are therefore often preferred for EEG.
- Gold-plated electrodes are also common but may drift more with very slow components.
- In EEG recordings all electrode impedances should be low and similar; below 5 kΩ is the usual target.