Filters, sampling and display settings
A filter makes a recording readable by attenuating certain frequencies. It has a cost: not only the noise but also the signal changes. Normal values are therefore valid only for the filter settings with which they were collected.
- Low-frequency filter (high-pass): removes slow drift. Raising it shrinks slow components and can make a wave look shorter.
- High-frequency filter (low-pass): reduces muscle and sharp noise. Lowering it spreads the peak, reduces its amplitude and makes it appear later.
- Notch filter: carves out the region around 50 Hz. It can blunt a spike; the real solution is to fix the electrode problem.
What does the cut-off frequency mean?
The cut-off frequency is not a sharp wall. At this point half of the signal power is lost (about a 30% fall in amplitude, i.e. −3 dB), and attenuation continues with a slope that depends on the filter order. Some signal outside the passband still gets through.
Sampling
A digital device measures the signal a certain number of times per second. The sampling rate must be at least twice the highest frequency in the signal; otherwise a high frequency appears as a false slow wave (aliasing). In practice, rates far above this limit are preferred.
Sensitivity and time base
Display settings do not change the signal but they do change perception. Low sensitivity hides small waves; high sensitivity makes normal activity overflow. Use a consistent gain when measuring onset: in conduction studies the apparent onset can shift visibly with gain.