Upper limb SEP

The median nerve is stimulated electrically at the wrist and the response is followed over the arm, neck and cortex. Intensity is enough to produce a small thumb movement, slightly above motor threshold.

Device settings
Stimulus
Median nerve, 0.2 ms square wave, above motor threshold
Rate
≈ 3-5 Hz (non-integer)
Filter
≈ 30 – 1000 Hz (generally up to 3000 Hz)
Repetitions
≥ 500, twice
Analysis time
≈ 40-50 ms
ErbN9C5-C7N13CortexN20040 ms
Three recording points in median SEP (negative up): Erb (N9), cervical (N13) and contralateral centro-parietal cortex (N20).
Waves and approximate sources
WaveRecording siteApproximate source
N9Erb's pointBrachial plexus
N13Over C5-C7Cervical dorsal horn, dorsal column entry
P14Scalp, relative to referenceCaudal medial lemniscus
N20Contralateral scalpPrimary sensory cortex

Central conduction time (N13-N20) is very valuable, as it excludes peripheral nerve and arm length effects. To see peripheral delay, first look at N9.

Finding → lesion site
FindingSite to consider
N9 delayed or absentPeripheral nerve or plexus
N9 normal, N13 absentRoot / posterior horn / dorsal column entry
N13 normal, N20 absentAbove cervical level, brainstem or thalamocortical pathway
N13-N20 prolongedSlowed central conduction

Coma prognosis

After cardiac arrest, once the effect of sedation and temperature management has passed, bilateral absence of N20 is among the strong indicators of poor prognosis. The peripheral response (N9, N13) must be intact, and the finding alone is not enough for decisions; multimodal assessment is needed.