Sympathetic skin response

After a sudden stimulus (electrical, deep breath), a voltage change is recorded on the palm and sole. This response reflects the function of the thin sympathetic C fibres that activate sweat glands and partly covers an area thick-fibre studies cannot see.

Device settings
Recording
Palm and sole, dorsum as reference
Filter
≈ 0.1-2000 Hz, low cut-off matters
Sweep
≈ 5-10 s
Latency (approximate)
Hand ≈ 1.4-1.6 s, foot ≈ 2 s
0246 sStimulusHandFoot
Sympathetic skin response in hand and foot. Latency is longer in the foot.

The response shrinks with repetition through habituation and falls with age. Complete absence of the response is considered significant; because amplitude variability is high, small differences are hard to interpret.